COMPOUND IDENTIFICATION
Compound name
- gamma-Phenyl-gamma-butyrolactone
CHROMATOGRAM
ANALYTICAL REPORT
|
Column Trade Name (Column size / Particle size) |
CHIRALPAK® IK (4.6 x 250 mm 5 µm) |
|---|---|
| Chromatographic Mode | Normal Phase |
| Mobile Phase Composition | n-hexane / 2-propanol = 93 / 7 |
| Flow rate (mL/min) | 1.0 | Temperature (℃) | 25 |
|---|---|---|---|
| Injection Volume (µL) | 2.0 | Sample Concentration |
1.5 mg/mL in Mobile phase |
| Detection | UV-VIS 204 nm | ||
| Elution Order | |||
| No. | Rt | K' |
|---|---|---|
| 1 | 34.1 | 10.36 |
| 2 | 36.6 | 11.21 |
| Alpha Value | Resolution | |
|---|---|---|
| 1 - 2 | 1.08 | 2.43 |