APPLICATION ID 1402

Ethiazide

COMPOUND IDENTIFICATION

Compound Name Ethiazide
Ethiazide

CHROMATOGRAM

Chromatogram

ANALYTICAL REPORT

Column Trade Name

Column Size / Particle Size

CHIRALPAK® ID

250 x 4.6mm / 5µm

Mobile Phase Composition

n-hexane / ethanol / diethylamine = 70 / 30 / 0.1

Chromatographic Mode

Normal Phase

Flow rate (ml/min)

1.0

Injection amount (mg)

0.005

Temperature (°C)

40

Pressure (MPa)

Detection

UV 230 nm

Elution order

No.

Rt

K’

1

6

1

2

8

1.7

Alpha Value

Resolution

1 - 2

1.67

3.7